Pedicle screws are often used to secure surgically implanted hardware to the spine in patients with spinal disease or spinal trauma. In some cases, these screws loosen over time, leading to spinal instability and consequent pain. This is a common complication of spine surgery. One reason suggested for pedicle screw loosening is implant-associated infection, but until now there has been little clinical evidence to support this theory.
In the article "High frequency of low-virulent microorganisms detected by sonication of pedicle screws: a potential cause for implant failure," published today in the Journal of Neurosurgery: Spine, Vincent Prinz, MD, and colleagues recount their detection of microbial colonization of surgical implant hardware in patients with no apparent clinical signs of infection.
The study focused on 82 patients who underwent surgery during which a spinal hardware implant had to be removed. None of these patients had clinical or laboratory evidence of active infection, and there was no obvious sign of infection during surgery. Following surgery, sonication was used to detect the presence of microbial biofilms on the extracted hardware. This application of sound wave energy breaks apart any sticky biofilm covering surgical implants, separating microorganisms that inhabit it. The microorganisms that had colonized the implants were then examined to specify what types were present.
The authors found evidence of pedicle screw loosening in 54 of the 82 patients; the other 28 patients formed a control group. Sonication verified microbial colonization of extracted hardware with one or more bacteria in 22 (40.7%) of the 54 patients with pedicle screw loosening. Sonication found no microbial colonization in hardware removed from the 28 patients in whom there was no pedicle loosening.
The most commonly found microorganisms were identified as coagulase-negative staphylococci (specifically Staphylococcus epidermis, S. hominis, S. lugdunensis, S. haemolyticus, S. cohnii, and S. saprophyticus). These commensal bacteria are part of the normal flora covering human skin and mucous membranes, where they rarely cause harm to their hosts. The second most commonly found microorganism was Cutibacterium acnes, which, as its name indicates, causes acne and is also commonly found on human skin. All these bacteria carry a low level of virulence. When they aggregate on medical devices implanted within a patient, they form biofilms and can produce infections that show no clinical symptoms and are not easily detected.
The authors examined patient files to find the durations of the initial surgeries in which spinal hardware had first been implanted in the spine. They found a significant difference between the length of surgery in patients in whom microbial colonization of hardware was detected (mean 288 ± 147 minutes) and patients in whom no microbial colonization was found (mean 201 ± 103 minutes).
Based on the relationship between pedicle screw loosening and microbial colonization, the authors suggest that patients in whom infected hardware is replaced would benefit from "systemic biofilm-active antibiotic therapy to prevent new screw loosening." "Alternatively," they add, "local antibiotics may be applied, such as the antimicrobial coating of screws or application of local antimicrobial hydrogel or bone substitutes."
The authors summarize their findings in four take-away points:
- Possible low-virulent implant infection was not revealed by standard preoperative tests.
- There was a significant association between pedicle screw loosening and the presence of low-virulent pathogens on spinal implants.
- Coagulase-negative staphylococci were the most commonly found pathogens colonizing the implants.
- There was a significant association between long surgery times and a higher rate of implant infection.
When asked about the importance of the study, Vajkoczy said, "The results of our study will add to a paradigm shift in spine surgery, if not more generally, to the common handling of medical hardware implants. Our increasing awareness of slow virulent infections as a cause of late hardware failure has changed our concept of prevention and treating this unfortunate complication of spine surgery. Today, strategies are needed to prevent perioperative implant colonization, disrupt biofilms, and actively search for implant infections by using sonication and advanced microbiological analysis at the time of re-operation. The good news is that infection is a complication that can be treated successfully, if properly diagnosed."
Reference: Prinz V, Bayerl S, Renz N, Trampuz A, Czabanka M, Woitzik J, Vajkoczy P, Finger T: High frequency of low-virulent microorganisms detected by sonication of pedicle screws: a potential cause for implant failure. J Neurosurg Spine, published ahead of print May 28, 2019. DOI: 10.3171/2019.1.SPINE181025.
Source: Journal of Neurosurgery
Comprehensive Strategies in Wound Care: Insights From Madhavi Ponnapalli, MD
November 22nd 2024Madhavi Ponnapalli, MD, discusses effective wound care strategies, including debridement techniques, offloading modalities, appropriate dressing selection, compression therapy, and nutritional needs for optimal healing outcomes.
The Leapfrog Group and the Positive Effect on Hospital Hand Hygiene
November 21st 2024The Leapfrog Group enhances hospital safety by publicizing hand hygiene performance, improving patient safety outcomes, and significantly reducing health care-associated infections through transparent standards and monitoring initiatives.
Why Clinical Expertise Is the Cornerstone to Your Most Profitable Business Line
November 14th 2024Perioperative nurses bring vital skills in patient safety, infection control, and quality improvement. They enhance surgical outcomes and support health care systems during complex, high-risk procedures.
Strengthening Defenses: Integrating Infection Control With Antimicrobial Stewardship
October 11th 2024Use this handout to explain the basics of why infection prevention and control and antimicrobial stewardship are essential and how the 2 fields must have a unified approach to patient and staff safety
Blood Product Overtransfusion Is a Global Issue: Here Are 5 Reasons the Practice Must Change
October 9th 2024If a patient receives treatment or therapy that they do not need, it can cause unnecessary harm. This is true for medications, surgeries, and medical procedures, especially blood transfusions.