The Sterilization Quality Assurance Process

Article

The Sterilization Quality Assurance Process

By Charles A. Hughes

Proper sterilization of instruments and materials is a critical aspect of infection control. Historically, healthcare professionals have sought a means to effectively and reliably manage sterilization by monitoring the parameters needed for sterilization (time, temperature and sterilant contact). Since microorganisms cannot be seen with the naked eye, the major difficulty in sterilization is determining when an item is considered sterile. In earlier days, a raw potato was sometimes placed in the sterilizer along with the load being processed and afterwards examined to see if it was cooked. If it was, then the load was considered sterile. As sterilization processes became more sophisticated, more scientific monitoring practices replaced this rather crude method. Today, an overall sterility assurance program involves the use of the following management tools:

Mechanical/Electronic Controls

Mechanical or electronic controls are the easiest for sterile processing personnel to use and interpret, and they provide the earliest indication of unsatisfactory sterilization conditions. While most sterilizers come with time, temperature and pressure gauges, these readings are limited as they only tell us what is happening at the probe site (usually in the drain) and not inside the packs being processed. The Association for the Advancement of Medical Instrumentation (AAMI) guidelines state that sterile processing personnel should review and record these readings after each and every sterilization cycle. Most older sterilizers use analog (mechanical) controls, while newer sterilizers usually have digital (electronic) controls. Compared to analog controls, digital controls have proven to be much more reliable; therefore, you should upgrade where possible. There are also new technology digital monitoring systems available that monitor all the parameters of the sterilization process through a real-time system with adjustable alarms to alert the user to any changes to the vessel, pumps, valves or temperature variances. These systems graph the physical parameters of the sterilization process that can be visually observed by the user.

Chemical Indicators/Integrators

Experienced practitioners use chemical indicators (CIs) as an adjunct to biological indicators in an effort to achieve a system of checks and balances. AAMI guidelines state that an external and internal CI should be used with every pack. The FDA classifies CIs as Class II medical devices requiring standards; AAMI, in "under Sterilization of healthcare products - Chemical indicators - Part 1: General requirements, Part 4.5," has established the following classifications.

1. Class I - Process indicators are for external use to show "processed" vs. "unprocessed" items and can usually be found in the form of tape or labels, as well as printed directly on some sterilization packaging.

2. Class II - Bowie-Dick indicators are used to show the pass/fail in steam vacuum-type sterilizers with the presence of air. Quality assurance is attained by performing an air removal test with an indicator sheet placed inside a towel pack (AAMI specified) or with a commercially prepared test pack that already includes an indicator sheet. AAMI standards state the Bowie-Dick test must be carried out each day the vacuum steam sterilizer is used, in an empty chamber.

3. Class III - Temperature-specific indicators are designed to reveal the attainment of specific minimum temperature at that location within the sterilizer chamber or load. This type is not sensitive to other sterilization parameters, such as time or sterilant.

4. Class IV - Multi-parameter indicators are much more accurate by design, as they provide an integrated response to all the parameters needed to achieve sterilization, by either a specified color change or migration along a wick to indicate pass or fail.

5. Class V - Integrating indicators are designed to react to all critical parameters over a specified range of sterilization cycles. The stated values are those required to achieve a stated inactivation by referring to a stated test organism with stated D and, if applicable, Z values.

Biological Indicators

The third tool used to manage sterilization involves the use of bacterial spores. Biological Indicators (BIs) are considered the highest level of sterility assurance because they actually test the sterilizer's ability to kill specific strains of highly resistant organisms. Commercially available BIs are non-pathogenic (harmless) and are specific for the type of sterilization process being monitored. BIs are available in the following formats:

1. Spore strips are available in a variety of concentrations and spore populations, allowing the user to select a specific sterility assurance level (SAL). After processing, the spore strip envelope is opened and the BIs aseptically transferred into a tube of sterile culture media, then incubated for up to seven days at a preset temperature (Geobacillus stearothermophilus = 55-60 degrees C or Bacillus atrophaeus = 35-37 degrees C).

2. Self-contained vials are available as single or dual species for use with steam or ethylene oxide (EO) gas processes. A glass ampule of sterile media is surrounded by a plastic vial with a spore strip inside. After processing, the user activates the BI by crushing the vial (manually or with a device) to let the media come into contact with the spore strip, then incubating at the appropriate temperature. No media color change of a processed BI indicates negative (sterile) results, providing an unprocessed BI that is used as a control cultures positive. Incubation time varies from 24 hours to seven days depending on the manufacturer.

3. Sealed glass ampules are available with Geobacillus stearothermophilus spores for use with steam processes. This form has the spores and media mixed together; therefore, no vial crushing or activation is required after processing. No media color change or visible turbidity (spore growth) of a processed BI indicates negative (sterile) results, providing an unprocessed BI used as a control cultures positive. Spore growth will usually occur within 8 to 12 hours; however, 48 hours incubation is recommended. Spore growth is evidenced by a yellow color change in the media along with visible turbidity at the bottom of the vial.

4. Spore suspension is available in pharmaceutical vials directly 10mL from manufacturers and is used for cycle and/or direct inoculation of product for process validation. Geobacillus stearothermophilus, Bacillus atrophaeus and B. pumilus spores are commonly stocked by suppliers.

BI Test Packs

For consistency in testing, AAMI guidelines recommend placing the BI inside a test pack to properly challenge the sterilization cycle. Test packs can be made in-house by the user per AAMI guidelines or purchased pre-made from suppliers. Unlike the Bowie-Dick test pack, which is always run in an empty sterilizer, the BI test pack should be run in a fully loaded chamber along with other packs. Place the BI test pack in the area of the sterilizer that is least favorable to sterilization (normally in the front bottom section, near the drain). Depending on the sterilizer being tested (steam vs. EO gas) and the type of monitoring (routine vs. installation), pack assembly and testing procedures vary.

BI Testing Frequency

BI testing frequency varies among healthcare users and suppliers. For example, medical device manufacturers are required by the Food and Drug Administration to spore-test every load processed in steam or EO gas. A minimum of 10 BIs must be located throughout the load and it cannot be released until full incubation of the BIs along with an unprocessed control for seven days. Some hospitals in Europe do not test with BIs at all, while most U.S. hospitals follow AAMI guidelines that recommend daily testing of steam and every load of EO gas or any load that includes an implant.

Private offices and clinics seldom test more often than monthly, except in the dental market where the ADA recommends weekly testing and a number of states have mandated "weekly" testing of all sterilizers. If your facility is considering a new technology sterilization process, be sure to ask the manufacturer for detailed information regarding the type of BI to use, format, test pack design and frequency of testing. In some cases, you may be required to have product testing done to validate certain items and/or materials to make sure they are compatible with the new process. The manufacturer of the sterilizer, the manufacturer of the items, and/or an independent testing laboratory should be able to assist you.

Positive BIs

Regardless of the testing frequency you choose or the type of sterilizer being tested, if any of the processed BIs culture positive (show spore growth), then that means the sterilizer is suspect and the following actions should be taken:

1. Report results to supervisor

2. Recall and resterilize all "suspected non-sterile" items

3. Verify spore growth via gram stain

4. Determine cause of failure and correct

5. Retest sterilizer before using again

AAMI guidelines state: "Until the results of retesting are satisfactory, the performance of the sterilizer must be considered in question."

Sterilizer Maintenance

Each manufacturer provides users who purchase their sterilizers with a manual that includes comprehensive care and maintenance instructions. According to AAMI guidelines, all sterilization equipment must receive periodic and routine maintenance to ensure accurate and reliable operations. The following maintenance procedures are considered the fourth management tool:

1. Routine maintenance: daily inspections and cleaning in accordance with the manufacturer's recommendations. This generally includes the cleaning of gaskets, chart pens, chamber drain screens and internal/external surfaces.

2. Preventive maintenance: periodic preventive maintenance in accordance with the manufacturer's instructions.

3. Scheduled maintenance: this level of maintenance should include lubrication of appropriate parts and replacement of expendable parts, i.e., steam traps.

4. Calibration: periodic calibration of items such as pressure and temperature gauges, timers, recording and control devices must be carried out by qualified personnel as specified in the manufacturer's instruction manual.

Record Keeping

Record keeping is considered an absolute essential for any facility involved in the sterilization of instruments and materials for disbursement. In the event of a recall, there must be a system in place for finding and reprocessing the items in question. This is accomplished by keeping accurate records of each and every sterilizer's performance. Maintenance records must be kept showing such things as: model and serial number of each sterilizer, date of servicing, reason for the service request, what was done, what parts were used, etc. In addition, evidence is required to document that items processed did indeed get sterilized. While each facility may have varying methods of how to logically keep or file records of their sterilizers' performance, all sterilization records should include the following:

1. Date, sterilizer number and load number

2. Lot number, general contents of load

3. Exposure time, temperature and pressure

4. Operator's name or initials

5. Chemical indicator results

6. Biological indicator results

7. Bowie-Dick test results (steam vacuum)

Documentation assures the monitoring of the sterilization process is occurring, assures that the cycle parameters have been met, and establishes accountability. In addition, sterilization record keeping helps personnel determine whether or not a recall is necessary and the extent of the recall if so. By knowing the contents of each load that was processed and the lot number, sterile processing personnel can determine how critical the recall is, should the sterility of the load become suspect. From a legal point of view, if your facility was to be sued for malpractice concerning a nosocomial infection, sterilization records can document presumption of legitimacy and shift the burden of proof to the plaintiff.

As we've seen, sterilization of instruments and materials can and must be managed through the use of various tools. With the very real and growing concern over infectious hepatitis and AIDS, it is more important than ever that you select the proper sterilization process and document that sterilization has indeed taken place for each item you process.

Charles A. Hughes is the general manager for SPSmedical Supply Corp. which is a sterilization research, development and testing facility. A corporate member of AAMI and OSAP, Hughes sits on numerous sterilization committees and speaks on current sterilization issues.

Recent Videos
Meet Alexander Sundermann, DrPH, CIC, FAPIC.
Veterinary Infection Prevention
Andreea Capilna, MD, PhD
Meet the Infection Control Today Editorial Advisory Board Members: Priya Pandya-Orozco, DNP, MSN, RN, PHN, CIC.
Meet Infection Control Today's Editorial Board Member: Tommy Davis, PhD, ACHE, APIC, BLS
Meet Shannon Simmons, DHSc, MPH, CIC.
David Levine, PhD, DPT, MPH, FAPTA
Henry Spratt, Infection Control Today's Editorial Advisory Board member
Infection Control Today Topic of the Month: Mental Health
Infection Control Today Topic of the Month: Mental Health
Related Content