Dror Rosengarten, MD, of the Institute of Pulmonology at Hadassah-Hebrew University Medical Center in Jerusalem, and colleagues, reported on a cluster of pseudo-infections in patients in a bronchoscopy unit who had Burkholderia cepacia isolated from bronchoalveolar lavage fluid samples
Dror Rosengarten, MD, of the Institute of Pulmonology at Hadassah-Hebrew University Medical Center in Jerusalem, and colleagues, reported on a cluster of pseudo-infections in patients in a bronchoscopy unit who had Burkholderia cepacia isolated from bronchoalveolar lavage fluid samples.
The researchers report that in December 2008, bronchoalveolar lavage fluid samples obtained from three patients were positive for Burkholderia cepacia complex on culture. Samples obtained from bronchoscopes and rinse-water samples obtained from the washer-disinfector were found to be positive for B. cepacia complex. The researchers report that samples obtained from the reverse osmosis water at the inlet to the washer-disinfector yielded no growth on culture; however, those cultures obtained from the basin surface gew B. cepacia complex on culture. The researchers add that the reprocessing procedure was observed but no substantial breach in technique was found. Upon examination, it was found that the antibacterial filter on the water supply line was missing and that this was the probable cause of the cluster.
The researchers note, "Multiple avenues for contamination ... have been described, including ineffective bronchoscope cleaning, problems with disinfectants, and contaminated reprocessing equipment. Although the use of automated endoscope reprocessors has helped healthcare workers to maintain standards of disinfection, these reprocessors may also be the source of bronchoscope contamination. Although the insides of these devices are periodically disinfected, their water supply tanks, tubing and pumps are not always in contact with disinfectants, and these areas may serve as reservoirs for contaminating pathogens. Biofilm on the surfaces of the basin may protect organisms during disinfection cycles." They add further that despite detailed recommendations for preventing contamination during bronchoscope reprocessing, "it appears that this problem is underrecognized and underreported."
Reference: Rosengarten D, et al. Cluster of Pseudoinfections with Burkholderia cepacia Associated with a Contaminated WasherDisinfector in a Bronchoscopy Unit. Infect Control Hosp Epidemiol 2010;31:769-771.
Vet IP Roundtable 2: Infection Control and Biosecurity Challenges in Veterinary Care
March 31st 2025Veterinary IPs highlight critical gaps in cleaning protocols, training, and biosecurity, stressing the urgent need for standardized, animal-specific infection prevention practices across diverse care settings.
Invisible, Indispensable: The Vital Role of AHRQ in Infection Prevention
March 25th 2025With health care systems under strain and infection preventionists being laid off nationwide, a little-known federal agency stands as a last line of defense against preventable patient harm. Yet the Agency for Healthcare Research and Quality (AHRQ) is now facing devastating cuts—threatening decades of progress in patient safety.
The Sterile Processing Conference Survival Guide: How to Make the Most of Your Next Event
March 25th 2025From expert speakers to cutting-edge tools, sterile processing conferences, like the 2025 HSPA Annual Conference and the SoCal SPA's Spring Conference, offer unmatched opportunities to grow your skills, expand your network, and strengthen your department's infection prevention game.
Redefining Material Compatibility in Sterilization: Insights From AAMI TIR17:2024
March 24th 2025AAMI TIR17:2024 provides updated, evidence-based guidance on material compatibility with sterilization modalities. It offers essential insights for medical device design and ensures safety without compromising functionality.